Detail Kyoritsu KEW 1062 Digital Multimeter
High Accuracy, High Performance and Reliable Measurements
• 0.02% basic DC accuracy
• Large display with 50, 000 counts
• Dual display for double indication for AC and DC, V and Hz, etc.
• True-RMS Measurements AC and AC+ DC
• Wide AC Frequency bandwidth from 10Hz to 100kHz *
• True-RMS or MEAN value detection mode can be selected *
• Fast Peak Hold response time of 250µ s
• Low-pass filter for motor drive measurements *
• Low Power © for resistance measurements on sensible electronic circuits by a low and safe test current *
• User calibration function
* Only for KEW 1062
Safety design for industrial use
• Complies with IEC 61010-1 CAT.III 1000V, CAT.IV 600V
• Terminal shutter to prevent incorrect test leads' insertion in current terminals
• Very wide operating temperature range from -20 to 55º C
Reliable support for data management
• Large data internal memory
• KEW 1062: 10, 000 data ( logging) , 100 data ( manual)
• Download data and Live Monitoring on a PC via the USB interface
( Option for USB Communication set)
Specifications
Detection mode: MEAN/ RMS ( switch)
DC V:
50.000/ 500.00/ 2400.0mV/ 5.0000/ 50.000/ 500.00/ 1000.0V
( Input impedance: Approx. 100M© [ 50/ 500/ 2400mV] , 10M© [ 5/ 50/ 500/ 1000V] )
| ± 0.02% rdg± 2dgt ( Basic accuracy)
AC V [ RMS] :
50.000/ 500.00mV/ 5.0000/ 50.000/ 500.00/ 1000.0V
( Input impedance: 11M© < 50pF [ 50/ 500mV/ 5V] , 10M© < 50pF [ 50/ 500/ 1000V] )
± 0.4% rdg± 30dgt ( Basic accuracy)
AC V [ MEAN] * :
50.000/ 500.00mV/ 5.0000/ 50.000/ 500.00/ 1000.0V
( Input impedance:
11M© < 50pF [ 50/ 500mV/ 5V] ,
10M© < 50pF [ 50/ 500/ 1000V] )
| ± 1% rdg± 30dgt ( Basic accuracy)
DCV+ ACV:
5.0000/ 50.000/ 500.00/ 1000.0V
( Input impedance: 11M© < 50pF [ 5V] , 10M© < 50pF [ 50/ 500/ 1000V] )
± 0.5% rdg± 10dgt ( Basic accuracy)
DC A: 500.00/ 5000.0µ A/ 50.000/ 500.00mA/ 5.0000/ 10.000A, | ± 0.2% rdg± 5dgt ( Basic accuracy)
AC A [ RMS] : 500.00/ 5000.0µ A/ 50.000/ 500.00mA/ 5.0000/ 10.000A, ± 0.75% rdg± 20dgt| ( Basic accuracy)
AC A [ MEAN] * :
500.00/ 5000.0µ A/ 50.000/ 500.00mA/ 5.0000/ 10.000A
| ± 1.5% rdg± 20dgt ( Basic accuracy)
DCA+ ACA: 500.00/ 5000.0µ A/ 50.000/ 500.00mA/ 5.0000/ 10.000A , ± 1% rdg± 10dgt ( Basic accuracy)
© : 500.00© / 5.0000/ 50.000/ 500.00k© / 5.0000/ 50.000M© , ± 0.05% rdg± 2dgt ( Basic accuracy)
LowPower-© * : 5.000/ 50.00/ 500.0k© / 5.000M © , | ± 0.2% rdg± 3dgt ( Basic accuracy)
Continuity buzzer: 500.0© ( The buzzer turns on for resistances lower than 100± 50© )
Diode test: 2.4000V| ± 1% rdg± 2dgt
Open curcuit voltage: < 5V ( Approx. 0.5mA Measuring Current)
Capacitance: 5.000/ 50.00/ 500.0nF/ 5.000/ 50.00/ 500.0µ F/ 5.000/ 50.00mF, | ± 1% rdg± 5dgt ( Basic accuracy)
Frequency: 2.000~ 9.999/ 9.00~ 99.99/ 90.0~ 999.9Hz/ 0.900~ 9.999/ 9.00~ 99.99kHz, | ± 0.02% rdg± 1dgt ( Basic accuracy)
DUTY: 10~ 90% | ± 1% rdg
Temperature: -200~ 1372º C| ± 1% rdg± 1.5º C, -328~ 2501.6º F| ± 1% rdg± 2.0º F ( with the use of K-type Temperature probe)
Measurement functions:
DC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature, Capacitor, Duty cycle ratio, Decibel ( dBV, dBm) , Continuity Check, Diode Test, LowPower-© *
Other functions:
Data Hold ( D/ H) , Auto Hold ( A/ H) , Peak Hold * ( P/ H) , Range Hold ( R/ H) , Maximum value ( MAX) , Minimum value ( MIN) , Average value ( AVG) , Zero Adjustment ( Capacitor, Resistance) , Relative values, Save to Memory, LCD backlight
Withstand voltage: 6, 880V/ 5 sec.
Applicable standards:
IEC 61010-1 CAT.IV 600V, CAT.III 1000V Pollution degree 2,
IEC 61010-031, IEC 61326-1 ( EMC)
Power source: R6P ( 1.5V) × 4 ( Auto power off: approx. 20 minutes)
Dimensions: 192( L) × 90( W) × 49( D) mm
Weight: Approx. 560g ( including batteries)
Accessories:
7220A ( Test leads) , R6P× 4, Instruction manual,
8926 ( Fuse [ 440mA/ 1000V] ) × 1 ( included)
8927 ( Fuse [ 10A/ 1000V] ) × 1 ( included)
* Only for KEW 1062