Detail Protek 6510 Analog Oscilloscope 100MHz
Features:
• Four traces may be simultaneously displayed in ALT-sweep
• Five vertical display modes: CH1, CH2, Dual, Add, Subtract
• Vertical mode triggering for displaying 2 unrelated signals
• Trigger hold-off circuit for synchronizing complex waveforms
• Sweep speeds to 2nS/ Div
• 12kV acceleration voltage for bright displays of low duty cycle signals
• Z axis input
• TV Synchronization for either line or field patterns
• 20MHz bandwidth limit switch for triggering signals in a noisy environment
Specifications
Vertical Amplifier ( CH 1 and CH 2)
Sensitivity: 5 mV/ Div to 5 V/ Div in 7 calibrated steps, ( X5 Mag 1mV/ Div to 1 V/ Div)
Accuracy: ± 3% ( ± 5% in X5 Mag)
Bandwidth ( -3dB) : DC to 100MHz
Rise Time: £ 3.5 ns, X5 Mag: 20MHz Bandwidth: £ 17 ns
Input Impedance: 1MW + 25 pF
Input Coupling: AC-Gnd-DC
Invert Channel: CH2
Display Modes: CH1, CH2, Add, Subtract, Dual ( Alt/ Chop)
Maximum Input Voltage: 400V ( DC or AC peak)
CH1 Signal Output: 20mV into 50W
Horizontal Sweep
• Timebase A - Sweep Time: 50ns/ Div to 0.2 s/ Div in 21 calibrated steps ( 1-2-5)
• Timebase A - X10 Mag: 5ns/ Div to 20ms/ Div
• Timebase A – Accuracy: ± 3% ( ± 5% in X10 Mag)
• Timebase A - Hold Off: Continuously adjustable to > 2X sweep time
• Timebase B - Sweep Time: 50ns/ Div to 10µ s/ Div in 8 steps ( 1-2-5)
• Timebase B – Accuracy: ± 3%
• Timebase B - Delay Time: Continuously variable from 0.2µ s to 2s
• Timebase B - Delay Modes: Continuous or triggered
• Timebase B - Sweep Displays: A, B, A alt B, B triggered, X-Y
X-Y Mode
CH1-X axis, CH2-Y axis Sensitivity: Same as CH1 vertical amplifier
Bandwidth ( X-axis) : DC to 2MHz
Phase Error: £ 3° DC to 10kHz
Triggering
Trigger Modes: Auto, Normal, TV-V, TV-L
Source: Int. CH2, line ext.
Coupling: AC
Slope: Positive or negative
Trigger Sensitivity:
• Int ( DC to 10MHz) : 0.5 Div
• Int ( 10MHz to 100MHz) : 1.5 Div
• Ext ( DC to 10MHz) : 200mV
• Ext ( 10MHz to 100MHz) : 500mV
• TV ( Int) : 1 Div Min
• TV ( Ext) : 1V P-P
External Trigger Input Impedance: 1MW + 20 pF
Maximum Input Volts: 400V DC or AC peak
Z Axis Modulation
Sensitivity : ± 5V P-P Intensity increases with a negative going voltage input
Bandwidth ( -3dB) : DC to 2MHz
Input Impedance: Approx. 2 KW
Calibrator Signal
Voltage: 0.5V ± 3% P-P 1kHz Square Wave
Duty Cycle: Approx. 50%
CRT
Type: 6 " rectangular CRT with internal graticule. ( 1 Div = 1cm)
Acceleration Voltage: 12 kV
General Specification
Power Requirements: AC Input Voltage: 100V, 120V, 220V, 240V selectable
Frequency: 50/ 60Hz
Power Consumption: 55W
Operating Temperature: 0 to 40° C at 35% to 85% relative humidity
Dimensions: 5.25 " H × 12.67 " W × 14.5 " D
Weight: 24 lbs.
Supplied Accessories: Manual, 1 AC Power cord, ( 2) X10 Probes
Tampilkan Lebih Banyak